Chapter 6 Normal Mode
37
6
4. Insert sample workpiece 2 to check the measured value.
Enter the value of sample workpiece 2 using the numeric keys, and accept the
value by pressing the [ENTER] button.
When you press the [OK] button, the calibration is performed and a message
regarding the calibration result is displayed.
Input Conditions
Workpiece 1 < Workpiece 2
0.5 Span 2
If the above conditions are not satisfied, the “Calibration Error” message is
displayed.
When the above conditions are satisfied, the “Calibration OK” message is
displayed.