Commands and Responses
Read with external input to the DRQ terminal (DRQ read)
You can send an input signal (by short-circuiting DRQ (terminal number 6) and SG (terminal number
1, 4, or 5)) from an external device such as a PLC instead of sending a command.
Response
Error response
*1 The data length is different depending on the data being read. (Up to 10 characters)
*2 SK-1000: ±**.*** (reading range is -99.999 to +99.999)
*3 Data length: 2 characters
On the SK-1000 Series, the judgment output (HI/LO/GO output) and the alarm output ON/OFF
are displayed with 2-digit number (ASCII characters). Convert the 2-digit number (ASCII
characters) read to a binary number and check the ON/OF status of each bit to check the
judgment output status.
When the read data is "05":
"05" is converted to the binary number "0101".
When N.O.
Bit
Judgment output
When N.O. When N.C.
0
0: HI judgment output OFF,
1: HI judgment output ON
0: HI judgment output ON,
1: HI judgment output OFF
1
0: LO judgment output OFF,
1: LO judgment output ON
0: LO judgment output ON,
1: LO judgment output OFF
2
0: GO judgment output OFF,
1: GO judgment output ON
0: GO judgment output ON,
1: GO judgment output OFF
3
0: Alarm output ON,
1: Alarm output OFF
0: Alarm output ON,
1: Alarm output OFF
2 ms min.
Input signal
ON
OFF
SG
RD
SD
SG
SG
DRQ
DR
CR LF
Data of the sensor
amplifier with ID:00*
2
,, ,,
Data of the sensor
amplifier with ID:01*
2
Data of the sensor
amplifier with the last ID No.
*
2
,,
Sensor amplifier with ID:00 Sensor amplifier with ID:01 Sensor amplifier with the last ID No.
Control
output
*
3
Control
output
*
3
Control
output
*
3
ER CR LF
,
DR
,
Error No.
Reference
Bit 3: Alarm output ON Bit 0: HI judgment output ON
Bit 2: GO judgment output ON Bit 1: LO judgment output OFF
Commands and Responses
Writing
External devices such as PLC's use the following communication commands to write data to
DL-RS1A.
Attempting to write data with "read-only" attribute results in a communication error and an
error response (error number: 22).
Write to the specified sensor amplifier (SW command)
Command
Response
Error response
Write to all sensor amplifiers (AW command)
Command
Response
Error response
Point
SW CR LFID No. Data No.
,,
Setting data*
,
SW CR LFID No. Data No.
,,
ER CR LF
,
SW
,
Error No.
AW CR LFData No.
,
Setting data*
,
AW CR LFData No.
,
ER CR LF
,
AW
,
Error No.